基于SLM的Ti-6Al-4V人工骨支架的结构设计
投稿时间:2015-07-08  修订日期:2015-08-09
中文关键词:  SLM  人工骨支架  力学性能
基金项目:国家自然科学基金项目(面上项目,重点项目,重大项目)
中文摘要:
      设计了6种Ti-6Al-4V人工骨支架,利用SLM(选择性激光熔化)进行制备。采用万能材料实验机结合高速摄像机分析支架的压缩过程,并利用Ansys进行分析。结果表明:支架的立柱直径对支架的力学性能参数影响较大,横梁直径对其影响较小;在相同高度下,支架横梁直径增大、层数增多或立柱直径增大,均使结构长径比减小,使结构稳固且支架承载能力提高;支架的力学性能应与周围骨组织相匹配,经过分析,3种结构的支架力学性能参数与人体股骨相近,故可作为股骨的人工骨支架。
Structure Design of Ti-6Al-4V Artificial Bone Scaffold based on SLM
英文关键词:SLM  artificial bone scaffold  mechanical property
英文摘要:
      Six kinds of Ti-6Al-4V artificial bone scaffold were designed and built with SLM (Selective Laser Melting) technology. Universal material testing machine and high-speed camera were used to analyze the compression process of scaffolds, and ANSYS was used to analyze the stress of scaffold. The results showed that the vertical column diameter had a great influence on the mechanical property parameters of scaffolds but the beam diameter had little influence on them, under the same height, the increasing of beam diameter and layer or support column diameter, which result in structure length-diameter ratio decrease, so that structure firm and load-carrying capacity were improved and the mechanical property should match the surrounding bone tissue. Through the analysis, mechanical property parameters of three kinds of scaffolds were close to human femur, so they can be used as the artificial bone scaffolds of femur.
作者单位E-mail
杨文静 新疆大学 机械工程学院 401238209@qq.com 
乌日开西.艾依提 新疆大学 机械工程学院 wurikaixi@sina.com 
王娟 新疆大学 机械工程学院  
滕勇 解放军兰州军区乌鲁木齐总医院全军骨科中心  
摘要点击次数: 476
全文下载次数: 18
查看全文  查看/发表评论  下载PDF阅读器
关闭