Abstract:A simultaneous determination of oxygen and hydrogen in titanium and its alloys on infrared absorption has been described,and the optimal conditions have been obtained.The relative standard deviations(n=6)are under 10%,and the recoveries are within the range of 95.6%~103.9%.Time for each determination is within 3 min.The proposed method has been used for conventional analysis with satisfactory results.